GE Measurement & Control SolutionsGE Measurement & Control Solutions

Home » CO2 » Dewpoint IR » Vaporstat 9002
E-mail Print

Vaporstat 9002 Infrared Dew Point Sensor

Vaporstat 9002 Infrared Dew Point Sensor

 

The Vaporstat™ is the first commercial infrared dew point sensor that directly measures absolute humidity by counting the number of water molecules in air. The sensor display provides a user selectable display of grains/lb of dry air or °F dew point. The Vaporstat ™ is designed for direct in-duct or wall mounting applications and is compatible with all of Telaire's accessory enclosures. The Vaporstat ™ provides a similar performance to a chilled mirror, but is a fraction of the cost and requires considerably less adjustment and maintenance.

 


Features

Infrared Dew Point Offers:

  • A Break-Through in Low Cost Reliable Dew Point Sensing.
  • Measure Moisture Independent of Temperature Interference.
  • Direct Measurement of Dew Point Eliminates Stacked Errors Resulting From RH and Temperature Calculation.
  • Unprecedented 15-Year Sensor Life.
  • Non Interactive Sensing Technology That Can't be Poisoned or Saturated.
  • Can Be Field Calibrated to an Easy-to-Use NIST Reference Standard in Less Than 15 Minutes.
  • Advanced technology developed with the assistance of the: U.S. DOE, Desiccant Technologies Program, Oak Ridge National Laboratories.

Applications

Controls Humidity, Ensures Comfort & Prevents Mold Growth In:

  • Schools
  • Offices
  • Supermarkets & Big Box Retail
  • Hotels
  • Restaurants
  • Museums/Art Galleries
  • Ice Arenas
  • Laboratories
  • Clean Rooms

Specifications

Measurement Method

Non-dispersive infrared, dual-channel, non-iInteractive,non-saturating

Sensor Output Units

°F dew point

Measurement Range

0°F to 80°F (-18°C to 26°C) dew point

Typical Dew Point Accuracy @ 77°F (25°C), 27°F to 80°F (-3°C to 26°C) Dew Point

As measured against a factory certified reference: 35.6°F (2°C) Dew Point 3.6°F (-15.77°C)

NIST Certified Calibration

The NIST certified calibration of the sensor can be restored or checked using a zero gas certified to have less than 0.05 grains/lb dry air moisture content. (calibration kit 2076)

Altitude Correction

User adjustable in 500 ft (152.4 m) increments using keypad.

Operating Temperature Range Room and Duct

32°F to 120°F (0°C to 50°C); Accessory enclosure 1551 required

Outdoor

-20°F to 120°F (-29°C to 50°C) when installed in 1551 enclosure

Input Power

  • 18 to 30 VAC, 50/60Hz (half wave rectified)
  • 1.75 VA average, 2.75 VA peak

Analog Outputs (available simultaneously)

  • 0 to 10 VDC (100 Ω output impedance)
  • 4 to 20 mA (RLmaximum_500 Ω)

Relay Output

  • Normally open and normally closed (SPDT)
  • Gold bifurcated, 2A maximum @ 24 V

On Board Keypad Adjustments

  • Select °F dew point output and display
  • Altitude Correction
  • Measurement range
  • Analog output range
  • Zero concentration calibration (with nitrogen)
  • Calibration to ambient air
  • Relay setpoint
  • Relay dead-band

Limited Warranty

  • 18 months (see warranty card for details)

Sensor Rated Life

  • 15 years

Installation

  • Wall
  • Duct with aspiration box (model 1508)
  • Outside air enclosure (model 1551)

Not recommended for supply air duct mount installation