ISOVOLT Titan E Series

Titan E stationary X-ray machine

The ISOVOLT Titan E stationary X-ray machine is based on the field-proven ISOVOLT Titan series, but offers significant improvements in ease of operation, controllability and on-board logic to allow greater operational productivity and reliability. The new machine will find increasing application in meeting the requirements of system integrators for automated X-ray solutions, as well as being used as a stand-alone X-ray machine for conventional X-ray radiography and radiography .

Contact Sales

 

 

 

 

 

 

Features

Features

  • Single or three phase mains available
  • High stability
  • High accuracy
  • Precise reproducibility
  • Short ramp-up time (< 1.5 sec)
  • Extremely low ripple
  • High dose rate output
  • Intelligent X-ray tube warm-up
  • Large LCD display
  • Clear text messages in up to 16 languages
  • Speed sensitive rotary control (kV, mA)
  • Rugged and reliable design
  • Proven reliability
  • Full on-site maintainability
  • Local and remote diagnostics
  • Systems Integration via standard RS 232C optional: RS 422 or RS 485 / Profibus (EN 50170) / Ethernet
  • Remote operation up to 100 m (300 ft.)

Benefits

  • Tube protection due to automatic warm-up procedures and extended warm-up modes to safeguard tube performance.
  • Smart and intuitive user interaction, with several integration possibilities facilitating higher productivity.
  • Excellent dose reproducability with extremely stable high voltage section for optimized exposures.
  • Convenient integration into several external platforms, such as automated testing machines, leveraging different interface features for device control, monitoring and visualization.
  • Excellent endurance and performance for permanent or intermittent operation.
  • Less intensive maintenance combined with easier serviceability reduces total cost of ownership.
  • Broad range of tubes,accessories and kits available.

Applications

  • Low energy high power applications - Composite materials, low density plastics. Thin wall light-weight designs.
  • High energy, low power applications - Thick Aluminium castings with different wall thicknesses Heavy steel- or gray casts with different wall thickness range.
  • High Energy, High Power applications - Industrial CT for cast alloy materials. Improved POD on high density and thick material through higher current capability.
  • Fine tuned applications for measurement and calibration - Perfect fit for a number of scientific or life science applications as reference and calibration equipment.
  • Applications in integrated solutions - Standard system interface to allow communication to OEM solutions and systems. Modular and compact design, performs in various settings and can be programmed externallly Robust design for 24/7 operation

Accessories

ImageName & DescriptionsDownload

X-ray Tube Housing

acc-titan-e-wl5000

Water Cooling Pump for operation of unipolar X-ray tubehousings

Type: Water-Air
Cooling Capacity: max. 5000 W
Ambient Temperature: max. 25°C

acrobat-reader SE

acrobat-reader RE

acc-titan-e-wlk10

Water Cooling Aggregate WLK 10 for operation of unipolar X-ray tubehousings

Type: Water-Air
Cooling Capacity: max. 900 W
Ambient Temperature: max. 40°C

acrobat-reader

acc-titan-e-wlk31

Water Cooling Aggregate WLK 31 for operation of unipolar X-ray tubehousings

Type: Water-Air
Cooling Capacity: max. 3000 W
Ambient Temperature: max. 40°C

acrobat-reader

acc-titan-e-ow4002

Oil Cooling Pump OW 4002 for operation of bipolar X-ray tubehousings

Type: Water-Oil
Cooling Capacity: max. 4200 W
Ambient Temperature: max. 40°C

acrobat-reader
acc-titan-e-ol4503

Oil Cooling Pump OL 4503 for operation of bipolar X-ray tubehousings

Type: Oil-Air
Cooling Capacity: max. 4500 W
Ambient Temperature: max. 32°C

acrobat-reader
acc-titan-e-olk50

Oil Cooling Aggregate OLK 50 for operation of bipolar X-ray tubehousings

Type: Oil-Air
Cooling Capacity: max. 4500 W
Ambient Temperature: max. 40°C

acrobat-reader

acc-titan-e-water-turbines

Water Turbine Flow Rate Monitor for monitoring the cooling water flow

The fail-safe water turbine flow rate monitor is used to monitor the cooling water flow when operating an ISOVOLT X-ray tube with coolant from the tap water supply. The flow rate monitor is factory set and ready for installation: Water flow rate disconnect at 4 l / min and reconnect at approx. 4.5 l / min.

acrobat-reader

acc-titan-e-flash-lamp

X-Ray Warning Flash Lamp with "Fail-Safe Circuit"

When X-rays are being generated from an X-ray tube, there must be a clear and safe indication of this hazardous condition. All Seifert controls are equipped with multiple warning facilities, including the "High Voltage ON" indication lamp on the control panel.

While other external warning lamps might fail due to a defective bulb or other reasons, this "High Voltage ON" lamp is connected in a fail-safe circuit causing the high voltage to be interrupted in the event of bulb burnout or other malfunction.

acrobat-reader

acc-titan-e-interlock-switchjpg

Three-Phase Interlock Switch

A primary interlock switch is used in radiation protection cabinets and rooms in conjunction with doors and flaps that can be opened under regular operative conditions, if the power supply of the X-ray radiating unit must be interrupted in the primary current circuit and if this disconnection through an interlock circuit must be physically linked to said doors and flaps.

acrobat-reader
acc-titan-e-tripod-stand

Tripod Stand

The tripod stand was developed to support the X-ray tubehousing at various inspection tasks (site inspection or similar). It allows a simple and quick alignment of the X-ray tube to various exposure positions.

acrobat-reader

acc-titan-e-inspection-stand

Pipe Inspection Stand

The pipe inspection stand was developed to meet the specific requirements in welding seam inspection under extreme conditions.

acrobat-reader

Specifications

Unipolar Systems

High Voltage Generator
Max. Output Voltage 160 kV 225 kV
Max. Output Current 45 mA 45 mA
Max. Output Power 4,5 kW, Limited by Tube Specification 4,5 kW, Limited by Tube Specification
High Voltage Ripple 5 V/mA (With High Voltage Cable 10 m), 40 kHz 5 V/mA (With High Voltage Cable 10 m), 40 kHz
Insulation Oil Oil
Housing Dimensions (Cathode) (W x D x H) 350 x 870 x 850 mm (13.8” x 34.3” x 33.5”) 350 x 870 x 850 mm (13.8” x 34.3” x 33.5”)
Weight (Cathode) 189 kg (417 lbs), Including Power Module 189 kg (417 lbs), Including Power Module
Tube Voltage
Preselection and Setting From 5 to 160 kV in 0.1 kV / 1 kV / 10 kV From 5 to 225 kV in 0.1 kV / 1 kV / 10 kV
Digital Display of Set and Actual Values Simultaneous 4 Digits Each Simultaneous 4 Digits Each
Display Resolution 0.1 kV 0.1 kV
Accuracy < ±1% < ±1%
Reproducibility ±0.01% at Constant Temperature Level ±0.01% at Constant Temperature Level
Temperature Drift < 65 ppm/°C < 65 ppm/°C
Tube Current
Preselection and Setting From 0.1 to 45 mA in 0.01 mA / 0.1 mA / 1 mA From 0.1 to 45 mA in 0.01 mA / 0.1 mA / 1 mA
Digital Display of Set and Actual Values Simultaneous 4 Digits Each Simultaneous 4 Digits Each
Display Resolution 0.1 mA / 0.01 mA 0.1 mA / 0.01 mA
Accuracy ± 1% ± 1%
Reproducibility ± 0.01% at Constant Temperature Level ± 0.01% at Constant Temperature Level
Temperature Drift < 65 ppm/°C < 65 ppm/°C
Exposure Time
Programmable Timer Non-Volatile Memory Non-Volatile Memory
Preselection and Setting From 0.1 to 99.9 Minutes in 0.1 Min. Increments or from 1 to 999 Sec. in 1 Sec. Increments or as direct Min./Sec. value (up to 99’59”) From 0.1 to 99.9 Minutes in 0.1 Min. Increments or from 1 to 999 Sec. in 1 Sec. Increments or as direct Min./Sec. value (up to 99’59”)
Digital Display of Set and Actual Values The Remaining Time Is Displayed, i.e. After a Mains Failure Exposure Can Be Continued Without any Time Error The Remaining Time Is Displayed, i.e. After a Mains Failure Exposure Can Be Continued Without any Time Error
Prewarning Time
Preselection and Setting Digital Setting From 2 to 250 Seconds or de-activated Digital Setting From 2 to 250 Seconds or de-activated
Programmed Mode
Number of Storable Programs 250 250
Warm-Up Automatic Intelligent Tube conditioning based on Real-time Clock Extended warm-up for special conditioning Automatic Intelligent Tube conditioning based on Real-time Clock Extended warm-up for special conditioning
X-ray Tube Setup 8 Tube selectable from a database of > 40 pre-programmed tubes 8 Tube selectable from a database of > 40 pre-programmed tubes
Operation History 512 Events (256 On/Off Events) 512 Events (256 On/Off Events)
Warm-up History 128 Events 128 Events
Control Module
Dimensions (W x D x H) 460 x 270 x 100 mm (18.1” x 10.6” x 3.9”) Built into Desk Housing 460 x 270 x 100 mm (18.1” x 10.6” x 3.9”) Built into Desk Housing
Weight 4.9 kg (10.8 lbs) Including Desk Housing 4.9 kg (10.8 lbs) Including Desk Housing
Connected Loads
Power connection 1N PE 230 V ± 10% 50/60 Hz 3N PE 400/230 V ±10%, 50/60 Hz, 3-Phase, Grounded Neutral TN-S or TN-C-S Mains (Star Connected System, Optional 3-Phase Isolation Transformer) 1N PE 230 V ± 10% 50/60 Hz 3N PE 400/230 V ±10%, 50/60 Hz, 3-Phase, Grounded Neutral TN-S or TN-C-S Mains (Star Connected System, Optional 3-Phase Isolation Transformer)
Grounding Separate Grounding for X-ray Tube and High Voltage Generator (Minimum 6 mm2) Separate Grounding for X-ray Tube and High Voltage Generator (Minimum 6 mm2)
Mains Fuses 63 A (1N PE) or 25 A (3N PE) Time-Delay Fuses, Customer-Supplied 63 A (1N PE) or 25 A (3N PE) Time-Delay Fuses, Customer-Supplied
Operating Temperature Range 0°C to +40°C 0°C to +40°C
Storage Temperature Range -30°C to +70°C -30°C to +70°C

Bipolar Systems

High Voltage Generator
Max. Output Voltage -160 kV (Cathode), +160 kV (Anode) -225 kV (Cathode), +225 kV (Anode)
Max. Output Current 45 mA 45 mA
Max. Output Power 4,5 kW (Cathode) 3 kW (Anode) Limited by Tube Specification 4,5 kW (Cathode) 3 kW (Anode) Limited by Tube Specification
High Voltage Ripple 10 V/mA (With High Voltage Cable 10 m), 40 kHz 10 V/mA (With High Voltage Cable 10 m), 40 kHz
Insulation Oil Oil
Operation History 512 Events (256 On / Off events) 512 Events (256 On / Off events)
Warm-up History 128 Events 128 Events
Housing Dimensions (Cathode) (W x D x H) 350 x 870 x 620 mm (13.8” x 34.3” x 24.4”) 350 x 870 x 620 mm (13.8” x 34.3” x 24.4”)
350 x 870 x 850 mm (13.8” x 34.3” x 33.5”) 350 x 870 x 850 mm (13.8” x 34.3” x 33.5”)
Weight (Anode) 123 kg (272 lbs) 123 kg (272 lbs)
Weight (Cathode) 189 kg (417 lbs), Including Power Module 189 kg (417 lbs), Including Power Module
Tube Voltage (Anode)
Preselection and Setting From 5 to 320 kV in 0.1 kV / 1 kV / 10 kV From 5 to 450 kV in 0.1 kV / 1 kV / 10 kV
Digital Display of Set and Actual Values Simultaneous 4 Digits Each Simultaneous 4 Digits Each
Display Resolution 0.1 kV 0.1 kV
Accuracy < ±1% < ±1%
Reproducibility ±0.01% at Constant Temperature Level ±0.01% at Constant Temperature Level
Temperature Drift < 65 ppm/°C < 65 ppm/°C
Tube Current
Preselection and Setting From 0.1 to 45 mA in 0.01 mA / 0.1 mA / 1 mA From 0.1 to 45 mA in 0.01 mA / 0.1 mA / 1 mA
Digital Display of Set and Actual Values Simultaneous 4 Digits Each Simultaneous 4 Digits Each
Display Resolution 0.1 mA / 0.01 mA 0.1 mA / 0.01 mA
Accuracy ± 1% ±1%
Reproducibility ± 0.01% at Constant Temperature Level ± 0.01% at Constant Temperature Level
Temperature Drift < 65 ppm/°C < 65 ppm/°C
Exposure Time
Programmable Timer Non-Volatile Memory Non-Volatile Memory
Preselection and Setting From 0.1 to 99.9 Minutes in 0.1 Min. Increments or From 1 to 999 Sec. in 1 Sec. Increments or as direct Min./Sec. value (up to 99’59”) From 0.1 to 99.9 Minutes in 0.1 Min. Increments or From 1 to 999 Sec. in 1 Sec. Increments or as direct Min./Sec. value (up to 99’59”)
Digital Display of Set and Actual Values The Remaining Time Is Displayed, i.e. After a Mains Failure Exposure Can Be Continued Without any Time Error The Remaining Time Is Displayed, i.e. After a Mains Failure Exposure Can Be Continued Without any Time Error
Prewarning Time
Preselection and Setting Digital Setting From 2 to 250 Seconds or de-activated Digital Setting From 2 to 250 Seconds or de-activated
Programmed Mode
Number of Storable Programs 250 250
Warm-Up Automatic Intelligent Tube conditioning based on Real-time Clock Extended warm-up for special conditioning Automatic Intelligent Tube conditioning based on Real-time Clock Extended warm-up for special conditioning
X-ray Tube Setup 8 Tube selectable from a database of > 40 pre-programmed tubes 8 Tube selectable from a database of > 40 pre-programmed tubes
Control Module
Dimensions (W x D x H) 460 x 270 x 100 mm (18.1” x 10.6” x 3.9”) Built into Desk Housing 460 x 270 x 100 mm (18.1” x 10.6” x 3.9”) Built into Desk Housing
Weight 4.9 kg (10.8 lbs) Including Desk Housing 4.9 kg (10.8 lbs) Including Desk Housing
Connected Loads
Power connection 1N PE 230 V ± 10% 50/60 Hz 3N PE 400/230 V ±10%, 50/60 Hz, 3-Phase, Grounded Neutral TN-S or TN-C-S Mains (Star Connected System, Optional 3-Phase Isolation Transformer) 1N PE 230 V ± 10% 50/60 Hz 3N PE 400/230 V ±10%, 50/60 Hz, 3-Phase, Grounded Neutral TN-S or TN-C-S Mains (Star Connected System, Optional 3-Phase Isolation Transformer)
Grounding Separate Grounding for X-ray Tube and High Voltage Generator (Minimum 6 mm2) Separate Grounding for X-ray Tube and High Voltage Generator (Minimum 6 mm2)
Mains Fuses 63 A (1N PE) or 25 A (3N PE) Time-Delay Fuses, Customer-Supplied 63 A (1N PE) or 25 A (3N PE) Time-Delay Fuses, Customer-Supplied
Operating Temperature Range 0°C to +40°C 0°C to +40°C
Storage Temperature Range -30°C to +70°C -30°C to +70°C

Download


Descriptions English Français Deutsch 日本語 RU Español
Column Stand Accessory Brochure acrobat-reader acrobat-reader
Corrosion Erosion Brochure acrobat-reader acrobat-reader
Industrial Radiography Image Forming Techniques acrobat-reader
Inspection Technologies Brochure acrobat-reader acrobat-reader acrobat-reader acrobat-reader
ISOVOLT 160 M2 / 0.4-0.4 Tube Brochure acrobat-reader acrobat-reader
ISOVOLT 160 M2 / 0.4-0.4HP Tube Brochure acrobat-reader acrobat-reader
ISOVOLT 160 M2 / 0.4-1.5 Tube Brochure acrobat-reader acrobat-reader
ISOVOLT 160 M2 / 0.4-3.0 Tube Brochure acrobat-reader acrobat-reader
ISOVOLT 160 MC 2 Tube Brochure acrobat-reader acrobat-reader
ISOVOLT 160 MM2 / 0.2 Tube Brochure acrobat-reader acrobat-reader
ISOVOLT 225 M2 / 0.4-1.5 Tube Brochure acrobat-reader acrobat-reader acrobat-reader
ISOVOLT 225 M2 / 0.4-3.0 Tube Brochure acrobat-reader acrobat-reader
ISOVOLT 225 MM2 / 02 Tube Brochure acrobat-reader acrobat-reader
ISOVOLT 320 / 7 Tube Brochure acrobat-reader acrobat-reader
ISOVOLT 320 M1/25-55 Tube Brochure acrobat-reader acrobat-reader
ISOVOLT 320 M2 / 0.4 - 1.0 HP acrobat-reader acrobat-reader
ISOVOLT 320 M2 / 4.5-13 Tube Brochure acrobat-reader acrobat-reader
ISOVOLT 420 / 5 Tube Brochure acrobat-reader acrobat-reader
ISOVOLT 420 M1/25-55 Tube Brochure acrobat-reader acrobat-reader
ISOVOLT 450 / 5 Tube Brochure acrobat-reader acrobat-reader
ISOVOLT 450 M1/10-35 Tube Brochure acrobat-reader acrobat-reader
ISOVOLT 450 M1/25-55 Tube Brochure acrobat-reader acrobat-reader
ISOVOLT 450 M2 / 10 Tube Brochure acrobat-reader acrobat-reader
ISOVOLT 450 M2/0.4-1.0HP Tube Brochure acrobat-reader acrobat-reader
ISOVOLT Titan E Brochure acrobat-reader acrobat-reader acrobat-reader acrobat-reader
MXR 320 / 23 Tube Brochure acrobat-reader
MXR 320 / 26 Tube Brochure acrobat-reader
Oil Cooling Aggregate OLK 50 Brochure acrobat-reader acrobat-reader
Oil Cooling Pump OL 4503 Brochure acrobat-reader acrobat-reader
Oil Cooling Pump OW 4002 Brochure acrobat-reader acrobat-reader
Pipe Inspection Stand for Tubehousing ISOVOLT 160 Brochure acrobat-reader acrobat-reader
Three-Phase Primary Interlock Switch Accessory Brochure acrobat-reader acrobat-reader
Tripod Stand with Clamp for Tubehousing 160 M Brochure acrobat-reader acrobat-reader
Water Cooling Aggregate WLK 10 Brochure acrobat-reader acrobat-reader
Water Cooling Aggregate WLK 31 Brochure acrobat-reader acrobat-reader
Water Cooling Pump WL 1001 Brochure acrobat-reader acrobat-reader
Water Cooling Pump WL 5000RE Brochure acrobat-reader acrobat-reader
Water Cooling Pump WL 5000SE Brochure acrobat-reader acrobat-reader
Water Turbine Flow Rate Monitor Brochure acrobat-reader acrobat-reader
X-Ray Warning Flash Lamp with \"FAIL-SAFE CIRCUIT\" Brochure acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader acrobat-reader
X-Ray Warning Flash Lamp with "FAIL-SAFE CIRCUIT" Brochure acrobat-reader


acrobat-reader